Week 1 · In-Person Competency · Foundations

Instrument processing

Dirty side to clean side, without ever crossing back. Your instructor watches you run the full cycle — ultrasonic, pouch, autoclave, indicator check, storage.

Practice instrument processing a minimum of three (3) times, or until competency is achieved — at least one attempt must be a teach-back demonstration to a peer, showing you can explain and perform the procedure.

Clean the instruments
1
Don appropriate PPE (utility gloves, mask, eyewear)
2
Transport contaminated instruments to the dirty side of the sterilization area
3
Place instruments into ultrasonic cleaner with proper solution
4
Run ultrasonic for recommended cycle time
5
Remove instruments and thoroughly rinse with water
6
Pat down instruments to dry
Sterilize & store
7
Package instruments in sterilization pouches with indicator
8
Seal and label pouches if required
9
Check autoclave water level and fill with distilled water if needed
10
Load pouches into autoclave properly (do not overload)
11
Select correct cycle to start autoclave
12
Verify sterilization indicators have changed appropriately
13
Store sterile instruments on the clean side of sterilization area

Dirty to clean — one direction, never back

Instruments only move forward through this line. Crossing back is the fastest way to fail this competency.

1
🦼
Dirty instruments
2
💧
Ultrasonic clean
3
🌬️
Rinse & dry
4
📦
Package & autoclave
5
✓
Sterile storage

What your instructor is watching for

Competency is achieved when the student consistently demonstrates efficiency, accuracy, and proper technique without instructor intervention.

✓Demonstrate proper flow from dirty to clean without cross-contamination
✓Correctly operate ultrasonic cleaner and follow all steps for instrument cleaning
✓Properly package and prepare instruments for sterilization
✓Accurately check and maintain autoclave water levels using distilled water
✓Safely and correctly load, run, and monitor the autoclave
✓Verify sterilization using indicator systems
✓Maintain organization, safety, and infection control standards throughout the process
The full cycle

Dirty side to clean side — ultrasonic through storage.

The full cycle, start to finish — dirty side to clean side, ultrasonic through storage.
📖

Review before class

The procedure above is what you will be checked off on. This section is the knowledge behind it — how you actually prove the sterilizer worked, which is the part written exams focus on.

1Three ways to check

Only one of them proves sterility.

Every sterilizer load gets monitored three different ways, and students routinely mix them up because all three involve something changing color. The distinction is what each one is actually capable of telling you.

  • ●Physical — the gauges and readouts on the machine. Tells you the cycle ran.
  • ●Chemical — indicators and integrators that change color. Tells you conditions were reached.
  • ●Biologic — spore tests. The only one that proves organisms were killed.
Hold on to this one sentence. Color change tells you the package went through a cycle. Only a spore test tells you the cycle worked.
2Process indicators

Outside the package. “I went through.”

These are the stripes or dots printed on the outside of a sterilization pouch — the ones that darken during the cycle.

Their job is narrow and useful: a glance at the color change confirms a package actually went through a sterilization cycle before you trust it as sterile. That is genuinely important — but it is all they do.

Check the pouch itself, too. A sterile package is only sterile while it is intact. If a pouch has a tear or a hole, the instruments must be repackaged and reprocessed — no matter how perfect the indicator looks.
3Process integrators

Inside the package. Three conditions at once.

Integrators go inside the instrument package, and they respond to a combination of pressure, temperature and time — which is why they are also called multiparameter indicators. All of the sterilization factors are integrated into one result.

They come as strips, tabs, or small tubes of colored liquid. The advantage of putting one inside every package is assurance of penetration — proof that the sterilizing agent actually reached the instruments in the middle of the pack, not just the outside of the pouch.

  • ●Indicator = outside the pack, single condition, “this was processed.”
  • ●Integrator = inside the pack, pressure + temperature + time, “conditions were met in here.”
Sterilization pouch before and after processing, with a process integrator strip shown separately
Indicator = outside the pack, this was processed. Integrator = inside the pack, the right conditions were met.
Check yourself
A pouch comes out of the autoclave. The stripe on the outside has turned dark. What have you confirmed?
Only that it was processed. An external process indicator distinguishes a processed package from an unprocessed one. It does not prove the right conditions were reached inside, and it certainly does not prove sterility. That takes an integrator inside the pack, and ultimately a spore test.
4The limitation

They do not indicate sterility.

This is the single most important line in the whole topic, and it is worded almost exactly this way on exams.

Process indicators and integrators give you immediate visual control of sterilizing conditions — you get an answer the moment the door opens, which is genuinely valuable. But they do not indicate sterility, and they are not a replacement for biologic monitoring.

A chemical reacting to heat and steam is not the same thing as a living organism dying. The only way to know spores were killed is to put spores in and see whether they survived.

5Biologic indicators

Spore tests — the actual proof.

Biologic indicators (BIs), also called spore tests, are vials or paper strips containing harmless bacterial spores that are highly resistant to heat. Spores are used precisely because they are the hardest thing to kill — if the cycle killed them, it killed everything else.

Three BIs are used in a test:

  • ✓Two go inside instrument packs, and the sterilizer runs under normal conditions.
  • ✓The third is set aside as a control — it never goes in the sterilizer.

Afterwards all three are cultured. If the two processed spores are killed — a negative culture — the cycle was successful. Culturing is usually handled through a mail-in monitoring service.

Why the third strip exists. The control is never sterilized, so it should grow. If it does not, the spores were dead before the test started and the whole result is meaningless. The control proves the test itself was valid.
Check yourself
You run a spore test. The two strips from inside the packs show no growth. The control strip also shows no growth. What does this mean?
The test is invalid. The control strip never enters the sterilizer, so it must grow — that is what proves the spores were alive and the culturing worked. If the control shows no growth, you have learned nothing about your sterilizer and the test has to be repeated.
6How often

At least weekly.

Biologic monitoring is the only way to determine whether sterilization has occurred and all bacteria and endospores have been killed — so it is not a once-in-a-while check.

The CDC, the American Dental Association and the Office of Safety and Asepsis Procedures Research Foundation (OSAP) all recommend at least weekly biologic testing of sterilization equipment.

Several states go further and require routine biologic checks at their own intervals — weekly, monthly, or cycle-specific rules such as spore testing every 40 hours of use, or every 30 days.

Know your state. Weekly is the professional recommendation, but the enforceable number is whatever your state dental board sets. When a state rule is stricter than the recommendation, the state rule is the one that governs.
7Before you start

The whole thing in 5 lines.

  • ✓Three kinds of monitoring: physical, chemical, biologic.
  • ✓Indicator = outside the pack. Integrator = inside, pressure + temperature + time.
  • ✓Neither one proves sterility — and neither replaces biologic monitoring.
  • ✓Spore tests are the only proof. Three BIs: two processed, one control that must grow.
  • ✓At least weekly — unless your state requires more.

Instructor Sign-Off

Student instructions: When you’re ready, tell your instructor. They’ll watch you perform it start to finish, then check you off on their own phone. The Instructor Sign-Off box on this page turns green when you pass. Not a pass yet? They’ll tell you what to fix, and you go again.